  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 83
Application settings
84 Nio MKII
at least one of the displays of the same type has already been
calibrated before using a sensor.
To equalize the displays:
1. Be sure the reference display is calibrated to the desired
luminance.
2. Set the Equalization option in the Application Settings dialog.
3. Close the Application Settings dialog.
4. Open the Display Settings dialog and calibrate display (2).
The display will be calibrated and additionally the display
luminance will be matched to the luminance of the reference
display (1).
MediCal Administrator tab
Figure 38: MediCal Administrator tab
Here you can connect to MediCal Administrator, if present.
To connect to MediCal Administrator, click the MediCal Administrator
button. As a result, the MediCal Administrator connection wizard starts.
Please follow the guidelines from the wizard.
The MediCal Administrator software is a hospital-based softcopy
image quality management system that keeps the consistency
data of every connected display system in a central database. Via