Application settings
Nio MKII 83
Application settings
Calibration tab
Figure 36: Calibration tab
Here you can check the Equalization option, allowing you to match all
the displays from a multi-head system to the luminance of the first
calibrated display of the system.
Suppose you work with a system containing 2 displays of the
same type, and you wish display (2) to have the same output
luminance as display (1). Therefore first calibrate display (1). Next,
calibrate display (2) while the Equalization option is checked.
The first calibrated display in the system is the reference display.
Figure 37: Dual-head equalization
The Equalization option will be available only if:
the system contains at least one display of the same type as the
one being calibrated
cd
(1) Reference display
(2) Calibrated display: Luminance automatically matched
to reference display