  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 93
Application settings
94
Application settings
Calibration tab
Figure 50: Calibration tab
Here you can check the Equalization option, allowing you to
match all the displays from a multi-head system to the
luminance of the first calibrated display of the system.
Suppose you work with a system containing 2 displays of the
same type, and you wish display (2) to have the same output
luminance as display (1). Therefore first calibrate display (1).
Next, calibrate display (2) while the Equalization option is
checked.
The first calibrated display in the system is the reference display.
Figure 51: Dual-head equalization
The Equalization option will be available only if: