Performing quality checks with test patterns8Performing quality checks with test patternsIntroductionYou can display test patterns to perform quality

Page 3

Performing quality checks with test patterns9To display a custom test pattern1. Open MediCal LE.2. Click on the Test Patterns... button.3. Double-clic

Page 4 - Installation

Performing quality checks with test patterns10Pre-defined test pattern descriptionSMPTEThe SMPTE pattern is a universal pattern with which you can che

Page 5 - Uninstalling MediCal LE

Performing quality checks with test patterns11Centered SquareThis pattern can be used to check the white luminance.Put the optical sensor on the white

Page 6 - Optical sensor connection

Performing quality checks with test patterns12UniformityThis pattern can be used to check if the white and black levels are uniform all over the scree

Page 7 - To connect the sensor:

Calibrating displays13Calibrating displaysNote:In the Options section, you can select which displays in the system need or need not to be controlled.

Page 8 - To use the sensor:

Calibrating displays147. Select what kind of display you are going to calibrate: LCD (flat panel display) or CRT (display with a picture tube).When do

Page 9 - Starting MediCal LE

Calibrating displays1510. When you have selected to calibrate an “LCD” display, check if geometry and phase are well-aligned.If they are well-aligned,

Page 10 - Introduction

Calibrating displays1615. When the calibration is finished successfully, the Calibration Successful message appears and a 'happy' sound is p

Page 11 - Pattern options

Options17Options1. Open MediCal LE.2. Click on the Options button.3. Select the display function for all displays of the system.We recommend to select

Page 12

www.barco.comB410015 - 02November 2004

Page 13

Working uncalibrated18Working uncalibratedYou can ignore the corrections you have performed on the display system with MediCal LE.Therefore, right-cli

Page 14 - Uniformity

Troubleshooting19TroubleshootingThe following problems may occur:The sensor is connected, but the system does not find itThis may be the case when you

Page 15 - Calibrating displays

Glossary20GlossaryFrequently-used terms:CalibrationIn many cases, calibration is the remedy for a display that is no longer conform to its target work

Page 16

Glossary21Gamma 2.2This is a gamma function with a gamma value of 2.2, which is approximately the natural gamma of most CRT displays.Imaging board hea

Page 17

About MediCal LE22About MediCal LEMediCal LE 1.03 versionCopyright © 2001-2004 BARCO nv. All rights reserved.More information? http://www.barco.com/me

Page 18

Warranty and License Agreement23Warranty and License AgreementSoftware License AgreementThis is a legal Agreement between you, the end user, and BARCO

Page 19

Warranty and License Agreement24TermThis license is effective until terminated. You may terminate it at any time by destroy-ing the Software and Produ

Page 20 - Working uncalibrated

Warranty and License Agreement25ExportYou acknowledge that the laws and regulations of the United States restrict the export and reexport of the Softw

Page 21 - Troubleshooting

Introduction1IntroductionMediCal LE is a user-friendly and efficient tool for Q/A check and conformance calibration of medical displays and display sy

Page 22 - Glossary

Installation2InstallationMediCal LE installationMinimum System requirements• Color or grayscale imaging board having gamma access• Operating system: W

Page 23 - Imaging board head

Installation35. Click on the Next> button.6. Read the License agreement on the screen very carefully. Scroll down to read the complete text.7. Clic

Page 24 - About MediCal LE

Installation4Optical sensor connectionSupported sensors:MediCal LE 1.03 supports 2 types of sensors.When sold as full package, MediCal LE 1.03 comes w

Page 25

Installation5Figure 1: Head adapter removal3. Pull the adapter from the sensor.To apply the head adapter to the sensor:1. Take the sensor with its “ey

Page 26

Installation6To use the sensor:The Barco LCD sensor, standard delivered with MediCal LE 1.03, is provided with a metal counterweight around the sensor

Page 27

Starting MediCal LE7Starting MediCal LEAfter installing MediCal LE, its icon appears in the System Tray in the Task Bar.The icon appears crossed out i